Jiuxin Medical Electric Co.Ltd |
|
|
|
|
|
|
|
|
Performance Specification And Tolerances Ventilator
Parameters Ranges Tolerances Controls Power ON/OFF Modes
IPPV
IPPV+SIGH
SIPPV
SIMV
MANUAL BREATH Tidal Volume 0~1200ml ¡À20% Breath Rate 5~60 bpm +/-20% PEEP 0~1KPa +/-20% %O2 45~90% Sigh Inspiratory Time 0.5~3sec +/-20% Trigger Level -0.5~1kpa +/-0.1kpa Over Pressure Relief less than 6kpa
Performance Specification And Tolerances
Ventilator
Parameters Ranges Tolerances Alarms and Indicators High Pressure Alarm Limit 2~6kpa
flash, red indicator +/-0.1kpa Low Pressure Alarm Limit 0. 2~2kpa
Red indicator +/-0.1kpa High Minute Volume Alarm Limit 10~30L
flash, Red indicator +/-20% Low Minute Volume Alarm Limit 1~10L
flash, Red indicator +/-20% High Rate Alarm Limit 30~59bpm
flash, Red indicator +/-20% Apnea interval 15, 30, 45, 60second flash, Red indicator +/-1second Alarm Silence 60 sec. max Power Red indicator Insp. phase Green indicator Exh. phase Green indicator T.VOL Indicator Green indicator MIN. VOL Indicator Green indicator PRESSURE Indicator Green indicator RATE Indicator Green indicator
|
|
|
|
|
|
|
|
|
|
|
|